On-line Total Organic Carbon Analyzer
On-line Total Organic Carbon Analyzer
+
  • On-line Total Organic Carbon Analyzer
  • On-line Total Organic Carbon Analyzer

On-line Total Organic Carbon Analyzer

TOC-800M-OL


The TOC-800M-OL online TOC analyzer is designed for medium/high TOC concentrations in water samples ranging from 0.050 ppm to 35,000.0 ppm, with high sensitivity, accuracy, and good stability.


Compact design, main components mounted on the main unit. Built-in gas generator; reagents required before measurement. Operated via a built-in Windows computer for measurement, calibration, printing, etc.


Key Features

1100℃ catalytic NDIR, 0.05-35000ppm, 85g/L salt

19" touch, printer, Windows, audit trail

Auto-sampler, precise pump, 3-stage condensate

Facebook
YouTube
LinkedIn
    • Brand: LabNav
    • Commodity name: On-line Total Organic Carbon Analyzer
    • Commodity ID: TOC-800M-OL
    • 产品卖点1: 1100℃ catalytic NDIR, 0.05-35000ppm, 85g/L salt
    • 产品卖点2: 19" touch, printer, Windows, audit trail
    • 产品卖点3: Auto-sampler, precise pump, 3-stage condensate

    Product Details

    Working Principle

    The sample is completely combusted at high temperature in the combustion furnace using a high-performance oxidation catalyst, decomposing it into carbon dioxide and water vapor. The water vapor is removed after cooling in a condenser, and the carbon dioxide is measured by a nondispersive infrared (NDIR) detector to determine the total carbon (TC) content of the sample. Separately, the inorganic carbon in the sample is decomposed into carbon dioxide and water vapor by an acid reagent; the water vapor is removed after cooling in a condenser, and the carbon dioxide is measured by an NDIR detector to determine the total inorganic carbon (TIC) content. The total organic carbon (TOC) is then calculated as TOC = TC – TIC.

     

    Applications

    Tap water, surface water, rivers, lakes Domestic sewage, industrial wastewater Chemical process water

     

    Measurement Software

    GMP-compliant: electronic signature, access control, data integrity (audit trail, non-deletable records, operation logging, method change logging, data backup).

     

    Features

    Large 19-inch display, built-in needle printer, unlimited data storage with password protection.

    Furnace up to 1100°C, choice of catalysts (CeO, Pt, CuO).

    High-reflection gold-plated gas cell, high-concentration IR source, and high-sensitivity detector for NDIR performance at the ppb level.

    Three-pass electronic condensation, automatic liquid injection with a precision electromagnetic pump, and non-manual injection.

    Real-time curve, multiple self-monitoring, overheat protection, and heated inorganic carbon reactor.

    Complies with ISO 8245, HJ 501-2009, and JJG 821-2005.

    Also see accessories.

     

    Specifications

    ParameterValue
    Detectable itemsTC, IC, TOC, NPOC
    TOC detection range0.050 – 35,000 mg/L
    Detection precision0.050 mg/L
    Furnace temperatureUp to 1100 ℃
    Accuracy error≤±5%
    Repeatability error≤3%
    Injection volume20–1000 μL
    Test time3–8 min
    Zero drift≤±2%/D
    Span drift≤±2%/D
    Salt tolerance85 g/L
    Acid reagentPhosphoric acid solution
    Ambient temperature0–60 °C (±5 °C/D)
    Relative humidity≤90%
    Power supply220 V AC ≤±10%, 50 Hz (reliable grounding)
    Power800 W
    Dimensions (W×D×H)60 cm × 60 cm × 130 cm
    Weight50 kg
    ConsumablesAlso see accessories.

     

    Main Components List

    ComponentDescriptionOrigin
    Solenoid valve200 kPa, bore 1.6 mmJapan
    Peristaltic pump60 ml/min, geared DC motorChina
    Temperature controllerAccuracy 0.1 ℃, contact life 1×10⁵, 0-50℃, ≤85% RHChina
    Micro pump20 μL per stroke, accuracy 1%, diaphragm PEEKUSA
    Dryer0-1000 ml/minUSA
    Combustion furnace680-1100 ℃Qikun
    NDIRRange: 0-5000 ppmGermany

     

    Ordering Information

    ModelOrder No.Electrical RequirementsConfiguration Description
    TOC-800M-OL106001014220V AC ≤ ±10%, 50 HzStandard: online main unit (built-in gas generator), combustion furnace, NDIR, software and tubing

     

    Instrument Configuration List

    No.ItemQty
    1Main unit1
    2Pyrolysis furnace1
    3Cerium oxide catalyst (40 g/pack)2
    4Quartz disc (15 g/pack)2
    5Quartz wool (0.5 g/pack)2
    6Halogen trap (15 g/pack)2
    7Quartz wool (2 g/pack)1
    8Injection septum4
    9Filter1
    10Filter membrane1
    11Oring6
    12Combustion tube2
    13Condenser coil1
    15Acid bottle1
    16PTFE adapter (6 to 3 mm)1
    17Injection nut (PEEK)1
    18Gas line nut (PTFE)1
    19Thumb screw (M5)2
    20Connecting tube (3.2×1.6 mm)8 m
    21Power cord1
    22Operating software1
    23User manual1
    24Certificate1
  • Consumables List

    ItemSpecificationUnit
    Quartz wool1-3 μmg
    Platinum-alumina catalystPt content 0.5%g
    Combustion tubepiece
    Halogen trap tubepiece