Total Organic Carbon Analyzer
Total Organic Carbon Analyzer
Total Organic Carbon Analyzer
+
  • Total Organic Carbon Analyzer
  • Total Organic Carbon Analyzer
  • Total Organic Carbon Analyzer

Total Organic Carbon Analyzer

TOC-600E


The TOC-600E Total Organic Carbon (TOC) Analyzer uses ultraviolet (UV) light to oxidize slowly flowing water samples. Organic compounds in the sample are oxidized to CO₂, and the CO₂ concentration is measured using conductivity detection technology. The TOC value of the water sample is obtained by comparing the conductivity before and after oxidation.

Key Features

Windows touch & audit trail

No catalyst/gas, 3min test

USP/ChP SST, autosampler

Facebook
YouTube
LinkedIn
    • Brand: LabNav
    • Commodity name: Total Organic Carbon Analyzer
    • Commodity ID: TOC-600E
    • 产品卖点1: Windows touch & audit trail
    • 产品卖点2: No catalyst/gas, 3min test
    • 产品卖点3: USP/ChP SST, autosampler

    Product Details

    Windows operating system—strong expandabilityData can be exported via USB.

    Optional autosamplers (24/30/70/81-position) save labor. 

     

    Working Principle

    Organic compounds in the sample are oxidized to carbon dioxide under UV light. The CO₂ concentration is measured using direct conductivity detection. The total organic carbon concentration is calculated by measuring the total inorganic carbon (TIC or IC) of the non-oxidized sample and the total carbon (TC) of the oxidized sample: TOC = TC – TIC

     

    Applications

    Pharmaceutical industry: Purified water, water for injection, sterile water for injection, and all deionized water.  
    Semiconductor industry: Electronic-grade process water.  
    Thermal power generation: Boiler feed water, boiler makeup water, steam condensate.  
    Laboratory water purifiers: Purified water, ultrapure water.

     

    Patents

    ZL 2014 2 0386326.8—A conductivity electrode with adjustable cell constant

    ZL 2014 2 0386169.0—A device for detecting total organic carbon in ultra-pure water

     

    Standards

    USP <643> Total Organic Carbon (TOC) method

    Chinese Pharmacopoeia 2020 edition, Volume IV, 0682 – Determination method for total organic carbon in water for injection

    GMP Annex – Computerized System Validation  
    21 CFR PART 11

     

    Features

    Built-in 8.4-inch computer screen operation – touch screen or mouse connection available 
    Upgradable software with audit trail function—no need for additional computer or module

    UV photocatalytic oxidation method – no catalyst or carrier gas required, easy maintenance.  
    Built-in full touch screen operation—simple, fast, reliable, and can connect to various printers.  
    Multi-level password authority management.  
    Automatic lock-on-leave protection.  
    Sample traceability – each sample has a unique number.  
    Complies with USP <643>, EP <2.2.44>, and Chinese Pharmacopoeia 2020, Vol. IV, Chapter 0682 "Total Organic Carbon Determination in Pharmaceutical Water."  
    Built-in system suitability test methods required by USP <643>, EP <2.2.44>, and Chinese Pharmacopoeia 0682.  
    Large memory automatically stores all test data—any test record can be queried and printed.  
    Fast analysis – single test time ≤ 3 minutes.  
    Small size, light weight, low energy consumption, and portable.

    Also see accessories.

     

    Specifications

    TOC Detection Range0.1 µg/L – 1500.0 µg/L
    Conductivity Detection Range0.055 – 6.000 µS/cm
    Analysis TimeMinimum 1 min interval (user-adjustable)
    Accuracy Error±4%
    Repeatability Error≤2 %
    Ambient Temperature0–60 ℃±5 ℃/D
    Relative Humidity≤90%
    Response TimeWithin 6 min
    Sample Temperature0–100 °C
    TOC Detection Precision0.1 µg/L
    Zero Drift±2 %/D
    Span Drift±2 %/D
    Power Supply220 ± 22 VAC, 50 ± 1 Hz
    Power Consumption60 W
    Dimensions (W×D×H)30 cm × 28 cm × 25 cm
    Weight8.6 kg
    ConsumablesAlso see accessories.

     

    Ordering Information

    ModelOrder No.Electrical RequirementsConfiguration Description
    TOC-600E106001002220 ± 22 VAC, 50 ± 1 HzUV lamp, peristaltic pump tube, calibration reagents
  • Consumables List
    NameSpecification/ModelReplacement Cycle
    UV lamp

    UV600

     

    5000 h
    Peristaltic pump tube1 year
    Calibration reagentsZero calibration water, sucrose stock solution, 1,4-benzoquinone stock solution (one bottle each)At calibration

DOWNLOAD