Total Organic Carbon Analyzer
+
  • Total Organic Carbon Analyzer

Total Organic Carbon Analyzer

TOC-700


The TOC-700 Total Organic Carbon (TOC) Analyzer uses UV + persulfate oxidation and NDIR detection principle. It can detect water samples with TOC concentrations ranging from 0.001 mg/L to 100.0 mg/L, featuring high sensitivity, accuracy, and good stability.

Key Features

UV/persulfate-NDIR, 0.001-100mg/L

10.4" touch, 4 modes, NDIR, precise pump

USP/ChP/21 CFR 11, autosampler & online

Facebook
YouTube
LinkedIn
    • Brand: LabNav
    • Commodity name: Total Organic Carbon Analyzer
    • Commodity ID: TOC-700
    • 产品卖点1: UV/persulfate-NDIR, 0.001-100mg/L
    • 产品卖点2: 10.4" touch, 4 modes, NDIR, precise pump
    • 产品卖点3: USP/ChP/21 CFR 11, autosampler & online

    Product Details

    Working Principle

    Difference method: The sample is fully oxidized to carbon dioxide and water by UV/persulfate oxidation. After condensation and dehydration, CO₂ is measured using a non-dispersive infrared (NDIR) detector to determine total carbon (TC) content. The inorganic carbon in the sample is decomposed into carbon dioxide and water by acid reagent. After condensation and dehydration, CO₂ is measured by NDIR to determine total inorganic carbon (TIC) content. TOC = TC – TIC.   

    Direct method: The inorganic carbon in the sample is first decomposed into carbon dioxide and water by acid reagent to remove total inorganic carbon (TIC). Then the organic carbon in the sample is decomposed into carbon dioxide and water by UV/persulfate oxidation. After condensation and dehydration, CO₂ is measured by NDIR to directly determine TOC content.

    10.4-inch display operation

    Simple and convenient consumables replacement

    Optional autosamplers

    (24 position / 30 position / 70 position / 81 position) save labor.

     

    Applications

    Deionized water samples such as water for injection and purified water   
    Pharmaceutical industry cleaning validation   
    Environmental water, such as tap water, surface water   
    Chemical process water (cleaning water, cooling water, recycled water, etc.)   
    Laboratory research

     

    Patents

    ZL 2014 2 0386326.8—A conductivity electrode with adjustable cell constant   
    ZL 2014 2 0386169.0—A device for detecting total organic carbon in ultrapure water

     

    Standards

    USP <643> Total Organic Carbon (TOC) method   
    Chinese Pharmacopoeia 2020, Vol. IV, 0682 – Determination of total organic carbon in water for injection   
    GMP Annex – Computerized System Validation   
    21 CFR PART 11

     

    Software Description

    The software meets GMP computerized system validation requirements: 
    Electronic Signature  
    Login via username and password, each unique. 
    Password length requirements enhance security. 
    The manual screen lock function ensures operational uniqueness and safety. 
    Access Control  
    Multi-level permission settings. 
    Administrators can assign different permissions to operators, ensuring personnel only have privileges commensurate with their job functions. 
    Data Integrity  
    Audit trail ffunction isenabled and cannot be turned off. 
    All test results are saved and cannot be modified or deleted. 
    All operations (any user entry) are recorded. 
    If a test method is modified, the changed parameters are recorded. 
    Data can be exported and backed up to protect stored data for future retrieval.

     

    Features

    Complies with EP, USP, and 2020 Chinese Pharmacopoeia requirements;   
    Four detection modes: TC, TIC, TOC, and NPOC, all in the same reactor.   
    High-reflectivity gold-plated gas cell, a high-concentration infrared light source, and a high-sensitivity infrared detector ensure excellent NDIR performance and sufficient sensitivity and accuracy for ppb-level measurements.   
    Precision electromagnetic metering pump ensures accuracy and stability of sample injection volume.   
    Dual-tube electronic condensation dehydration technology;   
    Real-time visible detection curve for better visualization;   
    Multiple real-time self-monitoring functions for temperature, pressure, and flow rate;   
    User definable calibration curve based on range, suitable for any water sample measurement;   
    Optional online module for continuous monitoring;   
    Optional autosampler for unattended operation, saving labor and time.

    Also see accessories.

     

    Specifications

    Analysis MethodUV/persulfate oxidation – NDIR detection principle
    TOC Detection Range0.001 mg/L – 100.0 mg/L
    Carrier GasHigh-purity nitrogen (CO₂-free air)
    OxidantSodium persulfate solution
    Analysis Time3–8 min
    Accuracy Error±5%
    Repeatability Error≤3 %
    Ambient Temperature0–40 ℃
    Relative Humidity≤ 85%
    Detectable ItemsTC, IC, TOC, NPOC
    Injection Volume50~6000μL
    Sample Temperature0-100℃
    Acid ReagentPhosphoric acid solution
    TOC Detection Precision0.001mg/L
    Zero Drift±2%/D
    Span Drift±2%/D
    Power Supply220±22V AC 50±1Hz
    Power Consumption150W
    Dimensions (W×D×H)42 cm × 36.5 cm × 38 cm
    Weight21 kg

     

    Ordering Information

    ModelOrder No.Electrical RequirementsConfiguration Description
    TOC-700106001008220 ± 22 VAC, 50 ± 1 HzStandard configuration